中文内容来自机器翻译,内容仅供参考,实际内容可参考原厂介绍或以合同为准。
英文原文:
Cost effective stationary biogas analyser designed for:
Discontinuous measurement, user settable up to 24 measurements per 24 hours
Use in the harsh industrial environment of combined heat and power (CHP) engine gas monitoring
Accurate measurements, incl. plug & play pre-calibrated sensors
Up to 2 sites monitoring (times sharing technique)
The analyser samples dry biogas, pressurized or low pressure gas and can be used for single or dual point sampling of:
Cogeneration heat and power engines (CHP)
Municipal or industrial waste water treatement sites
Small scale AD plants with dry fermentation
Landfill sites
Basic-Equipment:
Plug & play, pre-calibrated sensors
Safety in use with gas flow restrictor orifice at gas inlet and ventilation
Efficient sample gas preparation for fast and reliable measurements
Sampling from low suction up to high pressure gas
No dilution of the sample gas, neither use of compressed air is required
Discontinuous measurement, user settable up to 24 measurements per 24 hours
Up to 2 sites monitoring (time sharing technique) with only one analyzer
Data storage and event logging on SD card
IP 54 cabinet for use in harsh environment
Ready to run delivery, minimum installation work, low service downtime
May be installed outside or inside
Technical Data:
Measuring range:
CH4 0 ... 100% (infrared NDIR)
CO2 0 ... 100% (infrared NDIR)
O2 0 ... 25 % (electrochemical)
H2S 0 ... 2.000 ppm / overload up to 5.000 ppm (electrochemical)
Calculated values:
Calorific value 0 ... 50 MJ/m3; MJ/kg
Dimensions and other data:
Display 3,5" TFT colour display with backlight
Operating temperature +5 ... 45°C or -10 ... 45°C with cabinet heater
Power supply Universal 90 ... 240 Vac / 47 ... 63 Hz / 42W (150W with cabinet heater)
Protection IP 54
Dimensions 400x500x300 mm (WxHxD) wall or rack mounting
Weight approx. 14 kg